Filter for infusion set

Single use filtering device for use with any type of infusion set to prevent or protect from:

  • Bacteria.
  • Air inside the infusion set and syringe removed by air vent system.
  • Particles as fragments of glass or plastic.
 

Article extracted from "The specialist of anaesthesia and reanimation" March 1996 written by P. de Cillis

Glass ampoules are still the most commonly used recepticles in which to store injection solutions.
The process of opening ampoules prior to use can pose a risk of contamination caused by poor aseptic technique, and also, by tiny fragments of glass being deposited in the solution.
Published data eminating from various Italian groups has made reference to this problem.
A further risk may be posed by the fact that glass splinters derive mainly from the external face of ampoules, which can carry an invisible layer of microbial contamination. Additionally, shelving or other surfaces on which ampoules are stored can harbour bacteria. Ampoules containing injection solution are subjected to repeated handling prior to use and therefore the opportunity for contamination to occur, in the process of preparing parenteral therapy cannot be ignored.
Contamination, caused by glass splinters or microbial infestation may contribute either in combination, or separately, to granulomatous lesions found in various organs of pareterally treated patients.

Product specification
Air vent
0,02 µm hydrofobic PTFE
Housing material
Gamma resistant acrylic
Hydrophilic membrane
Polyethersulfone o Hydrophilic Cellulose Acetate
Hydrophilic membrane pore size
0,2 µm
Priming volume
< 0.4 ml overall (filter + tube) ~0.52 ml
Total lenght
~ 17 cm
Performance measurement
Flow rate @ 36" head height
3,0 ml/min
Hydrophilic membrane bubble point
a 35 psi (2,4 bar) minimum
Hydrophilic membrane burst pressure
75 psi (5,2 bar) minimum
Air vent minimum burst pressure
30 psi (2,4 bar) for 15 seconds
Filtration capacity
> 500 ml of normal saline

 


 

Istructions for use